PR electronics 5437B 2-wire HART 7 temperature transmitter

PR electronics 5437B 2-wire HART 7 temperature transmitter

Model: 5437B

Supplier: Products

Categories: PR electronics

Please contact us for more information.

PRODUCT DESCRIPTION

We promise 100% original brand new with 1-year warranty.

If you are interested in our products, please feel free to contact us!

-----------------------------------------------------------

HONGKONG XIEYUAN TECH CO., LIMITED

Add: 7-A16,Caishen Commercial Plaza,Hankou Railway Station,Wuhan, China

Contact: Ms. Lily

Email: lily@hkxytech.com

WhatsApp/Skype: +86 13789949182




PR-5437B




2-wire HART 7 temperature transmitter

 


  • RTD, TC, potentiometer, linear resistance and bipolar mV input

  • Single or true dual inputs with sensor redundancy and drift detection

  • Wide ambient operating temperature of -50 to +85°C

  • Total accuracy from 0.014%

  • 2.5 kVAC galvanic isolation

  • Full assessment to IEC61508 : 2010 up to SIL 2/3




Application

  • Temperature measurement of a wide range of TC and RTD types.

  • Conversion of wide span linear resistance and potentiometer inputs.

  • Conversion of bipolar mV signals to 4...20 mA.

  • Integration into asset management schemes.

  • Critical applications requiring superior accuracy and/or sensor redundancy and drift detection.





Technical characteristics

  • True dual input transmitter. High density 7-terminal design accepts the widest range of dual input combinations.

  • Sensor redundancy - output automatically switches to secondary sensor in event of primary sensor failure, maintaining uptime.

  • Sensor drift detection - alerts when sensor differential exceeds user-defined limits, for maintenance optimization.

  • Dynamic variable mapping for process data in addition to the primary variable e.g. dual input features such as average, differential and min./max. tracking.

  • Groundbreaking digital and analog signal accuracy over full input span and ambient conditions.

  • Extensive sensor matching including Callendar Van Dusen and custom linearizations.

  • Programmable input limits with runtime metering ensure maximum process traceability and sensor out of range protection.

  • IEC 61508 : 2010 full assessment up to SIL 3 together with enhanced EMC Functional Safety testing to IEC 61236-3-1.

  • Meets NAMUR NE21, NE43, NE44, NE89, NE95 and NE107 compliant diagnostics information.




Mounting / installation

  • For DIN form B sensor head mounting.

  • Configuration via standard HART communication interfaces or by PR 5909 Loop Link.

  • The 5437B can be mounted in zone 0, 1, 2 and zone 20, 21, 22 including M1.





Common specifications
Supply
Supply voltage7.5*...30** VDC
Internal power dissipation≤ 850 mW
Additional min. supply voltage when using test terminals0.8 V
Min. load resistance at >37 V supply(Vsupply – 37) / 23 mA
Isolation voltage
Isolation voltage, test / working2.5 kVAC / 42 VAC
Response time
Response time70 ms
Programmable damping0...60 s
Polarity protectionAll inputs and outputs
Warm-up time< 5 min.
Start-up time< 2.75 s
ProgrammingLoop Link & HART
Write protectionJumper or software
Signal / noise ratio> 60 dB
Long-term stability, better than±0.05% of span / year (±0.18% of span / 5 years)
Signal dynamics, input24 bit
Signal dynamics, output18 bit
Effect of supply voltage change< 0.005% of span / VDC
AccuracySee manual for details
EMC immunity influence< ±0.1% of span
Extended EMC immunity: NAMUR NE21, A criterion, burst< ±1%


Related Products